Unrecognized Failure of a Narrow Caliber Defibrillation Lead: The Role of Defibrillation Threshold Testing in Identifying an Unprotected Individual

Abstract: In this case report we describe a short circuit in the Riata 1570 defibrillator lead (Riata 1570, St. Jude Medical, St. Paul, MN, USA) that was unsuspected owing to normal lead parameters until defibrillator threshold testing at the time of elective generator change. On this occasion, the short circuit manifested as unsuccessful defibrillation of ventricular fibrillation with immediate battery depletion. This report adds weight to existing concerns over narrow caliber leads, it draws attention to the possibili… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
28
0

Year Published

2012
2012
2018
2018

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 31 publications
(28 citation statements)
references
References 5 publications
(10 reference statements)
0
28
0
Order By: Relevance
“…That these defects may not be detectable by routine monitoring is underscored by a case report showing that even intraoperative testing may fail to identify a shorted Riata lead. 11 Seven of the 22 Riata and Riata ST lead-related deaths were due to can abrasions that resulted in short circuiting between exposed high-voltage cables and the pulse generator. In these and other high-voltage failures, the pulse generator sustained electrical overstress and was unable to deliver further therapy.…”
Section: Discussionmentioning
confidence: 99%
“…That these defects may not be detectable by routine monitoring is underscored by a case report showing that even intraoperative testing may fail to identify a shorted Riata lead. 11 Seven of the 22 Riata and Riata ST lead-related deaths were due to can abrasions that resulted in short circuiting between exposed high-voltage cables and the pulse generator. In these and other high-voltage failures, the pulse generator sustained electrical overstress and was unable to deliver further therapy.…”
Section: Discussionmentioning
confidence: 99%
“…Moreover, in four cases there was no sign of externalization on fluoroscopy[14,16,18]. Our two patients had a 1571-8Fr and a 7000ST-7Fr, both dual-coil, both with EC.…”
Section: Discussionmentioning
confidence: 79%
“…The potential role of defibrillation testing in the management of Riata™ is currently unclear and has been poorly studied; only sparse reports are available in literature[14-19]. Some authors advocate it at time of pulse generator change[1,4,6,7] but patients at high risk could benefit from the test even before that time.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Serious adverse events, including deaths linked to Riata leads, have been reported 3, 4, 5, 6, 7. In these reports, the authors point to the risk of an HV short circuit caused by the inside-out abrasion underneath the shocking coils in the Riata lead family devices, though the incidence rate of internal abrasion short circuits underneath the SVC shock coil is quite low (0.06%) 8 .…”
Section: Discussionmentioning
confidence: 99%