Uma avaliação do rastreômetro, um novo equipamento para triagem populacional da pressão arterial elevada, em países em desenvolvimento

Abstract: objective: To test a simplified blood pressure device called Rastreometro that could be used by the Health Agents. Methods:The measurements made by the Rastreometro were compared with those by the auscultatory method, and were carried out in 268 patients, by two operators. The influence on the results of confounding variables such as age, gender, BMI, onfounding variables such as age, gender, BMI, arm length, upper arm circunference, skin colour and antihypertensive treatment were taken into consideration, as … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
4
0

Year Published

2012
2012
2012
2012

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(5 citation statements)
references
References 3 publications
1
4
0
Order By: Relevance
“…However, this situation can be explained in part by the fact of having been used a cuff smaller than indicated in the Rastreometer in 80.7% of the sample, which increased its sensitivity to some extent due to an overestimation of the arterial pressure 14 reading and did not influence the specificity. This finding confirms the finding of Forsvall 9 , which also has not identified differences in accuracy of the Rastreometer when using cuffs inadequate to arm circumference, however, in that study we used identical cuffs in the Rastreometer and a reference standard. Post-test probability -(%) 6,1 (2,5 -13,3) It is worth noting that the evaluation of the influence of arm circumference in the Rastreometer accuracy is impossible to achieve if using cuffs inadequate for arm circumference in the reference standard instrument, as the result obtained in this unit would not be consistent with the reality, but compatible with the Rastreometer measures.…”
Section: Bitencourt and Gottschall Evaluate The Accuracy Of Rastreometersupporting
confidence: 91%
See 4 more Smart Citations
“…However, this situation can be explained in part by the fact of having been used a cuff smaller than indicated in the Rastreometer in 80.7% of the sample, which increased its sensitivity to some extent due to an overestimation of the arterial pressure 14 reading and did not influence the specificity. This finding confirms the finding of Forsvall 9 , which also has not identified differences in accuracy of the Rastreometer when using cuffs inadequate to arm circumference, however, in that study we used identical cuffs in the Rastreometer and a reference standard. Post-test probability -(%) 6,1 (2,5 -13,3) It is worth noting that the evaluation of the influence of arm circumference in the Rastreometer accuracy is impossible to achieve if using cuffs inadequate for arm circumference in the reference standard instrument, as the result obtained in this unit would not be consistent with the reality, but compatible with the Rastreometer measures.…”
Section: Bitencourt and Gottschall Evaluate The Accuracy Of Rastreometersupporting
confidence: 91%
“…In the second study 9 the Rastreometer was tested on a sample of 268 patients, by two medical students, with results of 95.1% for sensitivity and 63.1% for specificity.…”