Ultraviolet Actinometry – Determination of the Incident Photon Flux and Quantum Yields for Photochemical Systems Using Low-Pressure and Ultraviolet Light-Emitting Diode Light Sources

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(3 citation statements)
references
References 7 publications
0
3
0
Order By: Relevance
“… Before and after the actinometry experiment, position the radiometer/spectroradiometer detector head so that the “calibration plane” (not necessarily the top of the detector head – this plane should be specified in the calibration certificate for the measurement device) is at the level of the top of the washer/ring mask or cardboard plate, and record the measurement device irradiance value. The incident photon flux [ q 0 (253.7) (einstein s –1 )] is then given by where E (253.7) is the irradiance (W m –2 ) at the center of the hole in the ring mask that is measured with the radiometer or spectroradiometer, the wavelength is according to the peak wavelength of the UV source [11] ,…”
Section: Methods Detailsmentioning
confidence: 99%
See 1 more Smart Citation
“… Before and after the actinometry experiment, position the radiometer/spectroradiometer detector head so that the “calibration plane” (not necessarily the top of the detector head – this plane should be specified in the calibration certificate for the measurement device) is at the level of the top of the washer/ring mask or cardboard plate, and record the measurement device irradiance value. The incident photon flux [ q 0 (253.7) (einstein s –1 )] is then given by where E (253.7) is the irradiance (W m –2 ) at the center of the hole in the ring mask that is measured with the radiometer or spectroradiometer, the wavelength is according to the peak wavelength of the UV source [11] ,…”
Section: Methods Detailsmentioning
confidence: 99%
“…Use of a monochromatic assumption for a UV-LED using the center wavelength has been compared to a process using numerical integration over the emission spectrum of the UV-LED. A simplification of the calculations has been examined by [11] using this protocol. No significant differences were observed between the monochromatic analysis and the polychromatic analysis, measured by the calibrated spectroradiometer and the ferrioxalate actinometer.…”
Section: Methods Detailsmentioning
confidence: 99%
“…However, they are also not broadly polychromatic as medium pressure mercury lamps which have emission lines throughout the 200-300 nm region of the spectrum and beyond. In a recent study, no significant differences were observed between monochromatic peak wavelength analysis and polychromatic analysis for spectroradiometer and ferrioxalate actinometer measurements (Pousty et al, 2022). However, testing using other criteria may show more polychromatic behavior, which would increase with the full width at half-maximum (FWHM) of the LED source tested, spectral response of the radiometer and actinometer, and could have varying impacts based on the action spectrum of microorganisms.…”
Section: Uvc Light Emitting Diodes As Disinfection Light Sourcesmentioning
confidence: 95%