Abstract:Device closure of secundum atrial septal defect (ASD) is the treatment of choice when anatomy is favourable. Amplatzer device has remained the gold standard for closure of ASD. Cobra deformity is a well-reported problem with devices. Recently, Tulip deformity has been reported in a single case. We report a series of cases where we noted Tulip deformity along with inability to retract the device in the sheath in Cera Lifetech devices. This resulted in prolongation of procedure, excessive fluoroscopic exposure a… Show more
Background:
Shape-memory abnormalities are seen in some nitinol atrial septal occluders. Variably described as cobra-head, tulip, and others, their incidence, mechanisms, clinical impact, and outcome have not been systematically analysed.
Methods:
We retrospectively reviewed all consecutive device closures in the last 6 years for deformations. Type and size of the occluder, deployment technique, size, and angulation/kinking of the delivery sheath were analysed. Procedural success, duration, and other complications were studied.
Results:
A total of 112 devices (11.8%) among 950 occluders used in 936 patients showed deformities. Fourteen of 936 received 2 devices. Deformities were transient and self-correcting in 40%. Multivariate analysis showed significant associations with oversized sheaths (p = 0.004), kinked/angulated sheaths (p < 0.001), special deployment techniques (p < 0.001), and twist in the device waist (p = 0.011). Despite more frequent deformities with Figulla (15.6%) and Amplatzer (13.9%) occluders than Cera occluders (6.6%) and larger devices (>24 mm – 14.6%) than smaller devices (less than or equal to 24 mm – 9.7%), they were not significant on multivariate analysis. In vivo manipulations corrected most deformities; nineteen needed in vitro reformations and four needed a change of device. Despite prolongation of the procedure, repeated attempts (mean 2.76 ± 1.7 attempts, with a range from 1 to 9 attempts), and supraventricular tachycardia in two patients, there were no serious adverse effects.
Conclusions:
Deformations were frequent in 11.8% of atrial septal occluders on a targeted search. Oversized and angulated/kinked sheaths, special techniques like pulmonary vein deployment and twist in device waist during procedure predisposed to deformities. While most deformities were corrected with manipulations, removal of the device was infrequently needed and change of device was rarely required. Long procedural time and multiple attempts for deployment did not affect procedural success.
Background:
Shape-memory abnormalities are seen in some nitinol atrial septal occluders. Variably described as cobra-head, tulip, and others, their incidence, mechanisms, clinical impact, and outcome have not been systematically analysed.
Methods:
We retrospectively reviewed all consecutive device closures in the last 6 years for deformations. Type and size of the occluder, deployment technique, size, and angulation/kinking of the delivery sheath were analysed. Procedural success, duration, and other complications were studied.
Results:
A total of 112 devices (11.8%) among 950 occluders used in 936 patients showed deformities. Fourteen of 936 received 2 devices. Deformities were transient and self-correcting in 40%. Multivariate analysis showed significant associations with oversized sheaths (p = 0.004), kinked/angulated sheaths (p < 0.001), special deployment techniques (p < 0.001), and twist in the device waist (p = 0.011). Despite more frequent deformities with Figulla (15.6%) and Amplatzer (13.9%) occluders than Cera occluders (6.6%) and larger devices (>24 mm – 14.6%) than smaller devices (less than or equal to 24 mm – 9.7%), they were not significant on multivariate analysis. In vivo manipulations corrected most deformities; nineteen needed in vitro reformations and four needed a change of device. Despite prolongation of the procedure, repeated attempts (mean 2.76 ± 1.7 attempts, with a range from 1 to 9 attempts), and supraventricular tachycardia in two patients, there were no serious adverse effects.
Conclusions:
Deformations were frequent in 11.8% of atrial septal occluders on a targeted search. Oversized and angulated/kinked sheaths, special techniques like pulmonary vein deployment and twist in device waist during procedure predisposed to deformities. While most deformities were corrected with manipulations, removal of the device was infrequently needed and change of device was rarely required. Long procedural time and multiple attempts for deployment did not affect procedural success.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.