Transmission Electron Microscopy (TEM) Specimen Preparation Technique using Focused Ion Beam (FIB): Application to Material Characterization of Chemical Vapor Deposition of Tungsten (W) and Tungsten Silicides (WSix)

Abstract: The specimen preparation technique using focused ion beam (FIB) to generate cross-sectional transmission electron microscopy (XTEM) samples of chemical vapor deposition (CVD) of Tungsten-plug (W-plug) and Tungsten Silicides (WSix) was studied. Using the combination method including two axes tilting[l], gas enhanced focused ion beam milling[2] and sacrificial metal coating on both sides of electron transmission membrane[3], it was possible to prepare a sample with minimal thickness (less than 1000 A) to get hig… Show more