Transcatheter aortic valve replacement with Lotus and Sapien 3 prosthetic valves: a systematic review and meta-analysis

Abstract: Background: Frequent occurrence of paravalvular leak (PVL) after transcatheter aortic valve replacement (TAVR) was the main concern with early-generation devices and focused technological improvements.Current systematic review and meta-analysis sought to compare outcomes of TAVR for severe native valve stenosis with next-generation devices: Lotus and Sapien 3. Methods: Electronic databases were screened for studies comparing outcomes of TAVR with Lotus and Sapien 3. In a random-effects meta-analysis, the poole… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
4
0
1

Year Published

2020
2020
2023
2023

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 7 publications
(6 citation statements)
references
References 70 publications
1
4
0
1
Order By: Relevance
“…More importantly, though, SCOPE I trial, by design, excluded over 300 patients with excessive calcification of aortic valve or left ventricular outflow tract (LVOT), which was not the case in remaining studies included in the current analysis. Presence of calcifications in both aortic annulus and LVOT could have accounted for much higher rates of PPI in the SAPIEN 3 arms across included PS-matched studies (average PPI incidence rate of 15.8%) as compared to SCOPE I trial with 9.3% rate, similarly to what has been already demonstrated for SAPIEN 3 in another meta-analysis by the same group [30].…”
Section: Discussionsupporting
confidence: 64%
See 1 more Smart Citation
“…More importantly, though, SCOPE I trial, by design, excluded over 300 patients with excessive calcification of aortic valve or left ventricular outflow tract (LVOT), which was not the case in remaining studies included in the current analysis. Presence of calcifications in both aortic annulus and LVOT could have accounted for much higher rates of PPI in the SAPIEN 3 arms across included PS-matched studies (average PPI incidence rate of 15.8%) as compared to SCOPE I trial with 9.3% rate, similarly to what has been already demonstrated for SAPIEN 3 in another meta-analysis by the same group [30].…”
Section: Discussionsupporting
confidence: 64%
“…Previous observational studies [15][16][17][18][19][20][21] and, among them, the SAVI-TF (Symetis ACURATE neo Valve Implantation Using Transfemoral Access) registry [29,30] reported on excellent short-term outcomes with low complications and, in particular, PPI rates in ACURATE neo valve attributable to the design of the prosthesis. The particularly low gradients also contributed to the similar or better rates of device success for ACURATE neo and SAPIEN 3 in propensity matched comparisons.…”
Section: Discussionmentioning
confidence: 99%
“…On review, we found 40 previous meta‐analyses discussing the risk factors of PPM implantation, however in light of the current evidence the applicability of those studies is limited. 92 , 95 , 96 , 97 , 98 , 99 , 100 , 101 , 102 , 103 , 104 , 105 , 106 , 107 , 108 , 109 , 110 , 111 , 112 , 113 , 114 , 115 , 116 , 117 , 118 , 119 , 120 , 121 , 122 , 123 , 124 , 125 , 126 , 127 , 128 , 129 , 130 , 131 , 132 Most of these meta‐analyses included a smaller number of previously published studies ranging from 4 to 41 articles, missing a large amount of contemporary data. The selection criteria and measured predictors were limited with respect to conduction abnormalities evaluated, indications for TAVR, and in some incidences inclusive of SAVR patients.…”
Section: Discussionmentioning
confidence: 99%
“…Consideration should also be given to the higher risk of some THV systems to induce atrioventricular conduction disturbance. However, it is important to realize that the choice of THV type/size and implantation depth is a complex interplay that will determine the final risk of PVL and conduction disturbance (34).…”
Section: Aortic Valve and Aortic Rootmentioning
confidence: 99%