“…In our practice with the first generation of triple-branched stent graft for arch repair in more than 100 patients with acute type A aortic dissection, we found that the diameters of 3 arch vessels could be easily matched by our prefabricated triple-branched stent grafts with the aid of banding at the bases of the arch vessels, and a difficult match occurred in the directions of 3 arch vessels, the distances between 2 neighboring arch vessels, and the tortuosity and diameter of the arch. 6 On the basis of these findings, a self-adaptive triple-branched stent graft, which could fit the different directions of 3 arch vessels, the various distances between 2 neighboring arch vessels, and the different tortuosity and diameter of the arch, was used in this study.…”