Timewalk correction for the Timepix3 chip obtained with real particle data

Abstract: In this work we have developed a prototype gaseous pixel detector by combining a micromegas grid with a Timepix3 chip for the readout. The micromegas foil supported by a matrix of pillars about 50 µm high was manually placed on top the chip. By placing a cathode foil above the chip an ionisation detector was created with a drift gap of 13.5 mm.The Timepix3 chip, thanks to the simultaneous measurement of the time-of-arrival (ToA) and charge via time-overthreshold (ToT) allows corrections to remaining timewalk e… Show more

Help me understand this report
View preprint versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
3
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 8 publications
(4 citation statements)
references
References 5 publications
1
3
0
Order By: Relevance
“…Figure 5 shows the simulated ToA for input charge values ranging from the used threshold level of 2.9 ke-to 10 ke-. The result is consistent with published results [7,8]. Also visible is the effect of the system noise of 80 e-ENC r.m.s, 7 e-r.m.s threshold noise and the long peaking time of the Timepix3 of ~35 ns for low charge inputs.…”
Section: Toa and Timewalksupporting
confidence: 92%
“…Figure 5 shows the simulated ToA for input charge values ranging from the used threshold level of 2.9 ke-to 10 ke-. The result is consistent with published results [7,8]. Also visible is the effect of the system noise of 80 e-ENC r.m.s, 7 e-r.m.s threshold noise and the long peaking time of the Timepix3 of ~35 ns for low charge inputs.…”
Section: Toa and Timewalksupporting
confidence: 92%
“…The position was defined as the center of mass of the individual blobs of triggered pixels. For determination of the ToF, the temporal information of the individual pixels was time-walk corrected by a pixel-based correlation curve between ToT and ToA [91][92][93]. The correlation curve was obtained from electron data for both detectors using the assumption that all electrons arrive at the same time within the resolution of the Timepix3 camera.…”
Section: Methodsmentioning
confidence: 99%
“…This effect is explained by the specific response characteristics of the in-pixel amplifiers and thresholding process. 49,50 The relative error introduced by these time-walk effects is not significant for the results in this work, and so no time-walk correction is performed. To illustrate the effect, a representative Ca + peak is shown in Figure S4 in the Supporting Information.…”
Section: The Journal Of Physical Chemistrymentioning
confidence: 99%