2014
DOI: 10.1016/j.chemosphere.2013.08.075
|View full text |Cite
|
Sign up to set email alerts
|

Time-integrated monitoring of polychlorinated dibenzo-p-dioxins and polychlorinated dibenzofurans (PCDD/Fs) in urban and industrial wastewaters using a ceramic toximeter and the CALUX bioassay

Abstract: The ceramic toximeter as a passive sampler in combination with the CALUX bioassay was utilized as a time-integrated pollution-assessment technique for dioxin-like PCDD/Fs in wastewaters. Toximeters filled with XCARB and enclosed in stainless steel cages were submerged in wastewater of Belbeis drainage-canal and EMAK paper-mill, located in Egypt, for 28 days. Two samplers were removed every week from each site. Extracts from toximeters, from bottom-sediments, and from paper-mill sludge were analyzed by the CALU… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
7
0

Year Published

2014
2014
2019
2019

Publication Types

Select...
4
2

Relationship

1
5

Authors

Journals

citations
Cited by 12 publications
(7 citation statements)
references
References 23 publications
0
7
0
Order By: Relevance
“…A value of m =2.0 was previously determined and used in field applications with ceramic-based passive samplers [12-14], but according to the results obtained on our testing solution of PCB-126, the value of m was set to 2.5. Since in this experiment only PCB-126 is present, a D e of 2.59 10 -7 will be used (see also Table 3 for diffusion coefficients of dl-PCBs).…”
Section: Resultsmentioning
confidence: 99%
See 4 more Smart Citations
“…A value of m =2.0 was previously determined and used in field applications with ceramic-based passive samplers [12-14], but according to the results obtained on our testing solution of PCB-126, the value of m was set to 2.5. Since in this experiment only PCB-126 is present, a D e of 2.59 10 -7 will be used (see also Table 3 for diffusion coefficients of dl-PCBs).…”
Section: Resultsmentioning
confidence: 99%
“…The ceramic cylinders were manufactured by ATECH Innovations (Germany) [12]. The XCARB (activated carbon 1% w/w on celite) was from Xenobiotic Detection Systems (USA).…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations