This is an open access article under the terms of the Creat ive Commo ns Attri bution-NonCo mmercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
AbstractBackground: Thrombin generation testing has been used to provide information on the coagulation phenotype of patients. The most used technique is the calibrated automated thrombogram (CAT) but it suffers from a lack of standardization, preventing its implementation in routine. The ST Genesia is a new analyzer designed to assess thrombin generation based on the same principle as the CAT. Unlike the CAT system, the ST Genesia is a benchtop, fully automated analyzer, able to perform the analyses individually and not by batch, with strict control of variables such as temperature and volumes, ensuring, theoretically, maximal reproducibility.
Objectives:This study aimed at assessing the performance of the STG-DrugScreen application on the ST Genesia analyzer. We also aimed at exploring stability of plasma samples after freezing and defining a reference normal range.
Results:Results demonstrated the excellent interexperiment precision of the ST Genesia and confirmed that the use of a reference plasma helps reducing the