2011
DOI: 10.1120/jacmp.v12i4.3581
|View full text |Cite
|
Sign up to set email alerts
|

Thermoluminescent and Monte Carlo dosimetry of IR06‐ brachytherapy source

Abstract: This work presents experimental dosimetry results for a new normalP103d brachytherapy seed, in accordance with the AAPM TG‐43U1 recommendation that all new low‐energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. Measurements were performed using TLD‐GR200A circular chip dosimeters using standard methods employing thermoluminescent dosimeters in a Perspex phantom. The Monte Carlo N‐particle (MCNP) code, version 5 was used to evalu… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
10
0

Year Published

2013
2013
2021
2021

Publication Types

Select...
5
2

Relationship

3
4

Authors

Journals

citations
Cited by 8 publications
(10 citation statements)
references
References 22 publications
0
10
0
Order By: Relevance
“…The responses of the TLD have to be corrected for background. This is done by subtracting the average response of background TLDs from the responses of all other TLDs in each measurement [25,32].…”
Section: Thermoluminescent Dosimetersmentioning
confidence: 99%
See 1 more Smart Citation
“…The responses of the TLD have to be corrected for background. This is done by subtracting the average response of background TLDs from the responses of all other TLDs in each measurement [25,32].…”
Section: Thermoluminescent Dosimetersmentioning
confidence: 99%
“…• Density: 1.19 g/cm 3 The design of the two phantoms to measure the radial dose and anisotropy functions are based on those of [25]. Figure 7 shows the phantom slab which is used for the experimental measurements of the brachytherapy seed radial dose function values.…”
Section: Phantomsmentioning
confidence: 99%
“…The resin beads are packed inside a titanium capsule of 4.7 mm length; 0.7 and 0.8 mm internal and external diameters, respectively; and 0.6 mm thick end caps and with an effective length of 3 mm. 103 -Pd radioactive material is absorbed uniformly in the resin bead volume [23]. Figure 1 shows a schematic diagram of the IR06-103 Pd seed.…”
Section: -Pd Source Descriptionmentioning
confidence: 99%
“…The composition of the Perspex taken to be H, 8%; C, 60%; O, 32%, with a density of 1.18 g/cm 3 . 13 This configuration enables us to verify the dose distributions around the applicator with a high spatial resolution. Fig.…”
Section: Ebt Gafchromic Film Dosimetrymentioning
confidence: 99%