2002
DOI: 10.1088/0031-9155/47/24/308
|View full text |Cite
|
Sign up to set email alerts
|

The use of diode array spectroradiometers for dosimetry in phototherapy

Abstract: An evaluation of two diode array radiometers, an UV spectroradiometer, Type SC-MP-A, from 4D Controls (Redruth, UK) and an USB2000-UV-VIS spectrometer from Ocean Optics (Duiven, NL), was carried out at the Photobiology Unit, University of Dundee. Three parameters of the instruments' performance were investigated, having been identified as the most likely sources of error in phototherapy dosimetry: (1) calibration, (2) stray light rejection, (3) angular response. An assessment was then made of the reliability o… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
11
0

Year Published

2004
2004
2016
2016

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 16 publications
(11 citation statements)
references
References 13 publications
0
11
0
Order By: Relevance
“…Each source was sampled using the USB2000 spectrometer with an integration time of 100 ms per scan to achieve maximum signal without saturation and an average of 200 scans to increase the signal to noise ratio (24) . Although the USB2000 has been reported to possess poor stray light rejection (25) , it was also shown that with adequate correction that the levels of stray light could be significantly reduced to less then 1 % at 250 nm (25) . A simple correction analogous to *Corresponding author: alanna.maguire@dit.ie background subtraction was employed in this study where the difference in signal between the blacked out pixels corresponding to 180 nm and at 250 nm, a wavelength at which no signal would be expected, was determined and subtracted at all wavelengths.…”
Section: Dosimetrymentioning
confidence: 99%
“…Each source was sampled using the USB2000 spectrometer with an integration time of 100 ms per scan to achieve maximum signal without saturation and an average of 200 scans to increase the signal to noise ratio (24) . Although the USB2000 has been reported to possess poor stray light rejection (25) , it was also shown that with adequate correction that the levels of stray light could be significantly reduced to less then 1 % at 250 nm (25) . A simple correction analogous to *Corresponding author: alanna.maguire@dit.ie background subtraction was employed in this study where the difference in signal between the blacked out pixels corresponding to 180 nm and at 250 nm, a wavelength at which no signal would be expected, was determined and subtracted at all wavelengths.…”
Section: Dosimetrymentioning
confidence: 99%
“…In most cases the Brewer spectrometers and other expensive spectrum scanning instruments are used. As an alternative, the advancement of technology has made available the compact and simple single-monochromator array spectrometers (JETI Technische Instrumente GmbH, 2005;Oliver and Moseley, 2002;Ylianttila et al, 2005). However, the limited dynamic range and intrinsic stray light problems complicate their use.…”
Section: Introductionmentioning
confidence: 99%
“…Evidence of reduced variability in UV radiometer calibrations is seen in intercomparisons from around Europe published in the 1980s, which showed a spread in measured calibration factors of UVA radiometers of 23%, while a more recent study demonstrated only a 6% spread . Finally, there have also been valuable instrumental advances, with the introduction of compact solid‐state spectroradiometers making it much easier than previously to capture spectral information from clinical UVR equipment …”
Section: Introductionmentioning
confidence: 99%