The targeting accuracy of a preclinical MRI‐guided focused ultrasound system

Abstract: The FUS system, as described, is capable of precise and accurate sonications. The largest source of error--misregistration of the coordinate systems of the scanner and ultrasound system--was addressed which reduced the error to 0.36 ± 0.14 mm, sufficient for many preclinical applications.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
13
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 22 publications
(13 citation statements)
references
References 25 publications
(25 reference statements)
0
13
0
Order By: Relevance
“…Due to the practical purposes, a small animal model would be most suitable for such a survival study. Dedicated preclinical focused ultrasound systems for small animals are available which can be used on 3T or 7T MRI systems, depending on local availability [ 36 38 ]. Treatment cells used in small animal studies should be reduced in size to a few mm since the sciatic nerves in rodents is on the order of 1 mm.…”
Section: Discussionmentioning
confidence: 99%
“…Due to the practical purposes, a small animal model would be most suitable for such a survival study. Dedicated preclinical focused ultrasound systems for small animals are available which can be used on 3T or 7T MRI systems, depending on local availability [ 36 38 ]. Treatment cells used in small animal studies should be reduced in size to a few mm since the sciatic nerves in rodents is on the order of 1 mm.…”
Section: Discussionmentioning
confidence: 99%
“…For the treatments, the animals were positioned supine above an MRI-compatible three-axis positioning system, operationally similar to that described by Ellens et al 23 , that would mechanically position the ultrasound focus by moving a single-element, fixed focus transducer. The experimental setup is illustrated in Fig.…”
Section: Methodsmentioning
confidence: 99%
“…The animals were placed supine on a sled above a 3‐axis positioning system, which was used to mechanically steer the transducer focus. The sled was designed to move between the positioning system, holding the US transducer, and the bore of a 7‐T MRI system (BioSpec 70/30 USR; Bruker, Billerica, MA) for treatment guidance.…”
Section: Methodsmentioning
confidence: 99%