1997
DOI: 10.1016/s0278-2391(97)90697-2
|View full text |Cite
|
Sign up to set email alerts
|

The pullout force for Mitek mini and micro suture anchor systems in human mandibular condyles

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
16
0

Year Published

2001
2001
2021
2021

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 32 publications
(16 citation statements)
references
References 5 publications
0
16
0
Order By: Relevance
“…Fields, Cardenas and Wolford [16] studied the pull-out strength of Mitek mini-anchors inserted in 20 human cadaver condyles, demonstrating an average pull-out load of 16.0 lb (range 8.5-28.4 lb). The pull-out strength of the orthodontic mini-implants range between 17.8 and 66.3 lb [17], and this strength appeared to be more than adequate to stabilize the disc to the anchor along with condylar movement.…”
Section: Discussionmentioning
confidence: 99%
“…Fields, Cardenas and Wolford [16] studied the pull-out strength of Mitek mini-anchors inserted in 20 human cadaver condyles, demonstrating an average pull-out load of 16.0 lb (range 8.5-28.4 lb). The pull-out strength of the orthodontic mini-implants range between 17.8 and 66.3 lb [17], and this strength appeared to be more than adequate to stabilize the disc to the anchor along with condylar movement.…”
Section: Discussionmentioning
confidence: 99%
“…Bone integration, holding intact the disc in proper place, and long-term stability of Mitek mini anchors in the TMJ is indebted to these two retention wings and sutures. [1112]…”
Section: Methodsmentioning
confidence: 99%
“…Laboratory and clinical research studies conducted at Baylor University Medical Center and Baylor College of Dentistry led to Food and Drug Administration approval of the Mitek mini anchor specifically for use in TMJ surgery in 1999 (6,17,(23)(24)(25)(26).…”
Section: Experience At Baylor University Medical Centermentioning
confidence: 99%
“…Fields, Cardenas, and Wolford evaluated the pullout strength of 20 Mitek mini anchors placed in human cadaver condyles (24). Pullout thresholds were reached when the cortical bone overlying the anchor gave way (18 condyles) or when the suture broke (2 condyles).…”
Section: Laboratory Researchmentioning
confidence: 99%