The metrology of electrical impedance at high frequency: a review

Abstract: The measurement of electrical impedance in the high-frequency range (loosely speaking, from 100 kHz to 100 MHz) is a common characterization technique in research and applications concerning condensed matter physics, engineering and biology. Commercial instruments for impedance measurement have specifications, including frequency range and base accuracy, continuously improving. The calibration of impedance instruments and standards in the high-frequency range represents a tough challenge for metrology laborato… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
26
0

Year Published

2011
2011
2019
2019

Publication Types

Select...
3
3
1

Relationship

0
7

Authors

Journals

citations
Cited by 43 publications
(33 citation statements)
references
References 78 publications
(33 reference statements)
0
26
0
Order By: Relevance
“…Bioelectrical impedance measurements have been widely used in the past 50 decades, giving rise to a great number of techniques and methods able to characterize solids, liquids and suspensions, and, lately, to biological tissues and fluids, either in vitro or in vivo [1], some of which having clinical applications [2]. Electric Impedance Spectroscopy, EIS, is an incoming technique that has been extensively used to characterize electrical properties of plant tissue [3], especially in what concerns to fruits and their respective quality [3] [5] [6], maturation process [4], and influence of temperature or other external agents [7], and to specific botanical tissues changes under controlled factors [8] [9] [10].…”
Section: Contextmentioning
confidence: 99%
See 2 more Smart Citations
“…Bioelectrical impedance measurements have been widely used in the past 50 decades, giving rise to a great number of techniques and methods able to characterize solids, liquids and suspensions, and, lately, to biological tissues and fluids, either in vitro or in vivo [1], some of which having clinical applications [2]. Electric Impedance Spectroscopy, EIS, is an incoming technique that has been extensively used to characterize electrical properties of plant tissue [3], especially in what concerns to fruits and their respective quality [3] [5] [6], maturation process [4], and influence of temperature or other external agents [7], and to specific botanical tissues changes under controlled factors [8] [9] [10].…”
Section: Contextmentioning
confidence: 99%
“…Nowadays, instruments having high accuracy, high resolution and frequency ranges extending from a few Hz to tens of MHz are commercially available [1]. Despites that, these solutions are quite expensive or, on the other hand, lack a true frequency bandwidth analysis, which is mandatory for the EIS technique.…”
Section: Contextmentioning
confidence: 99%
See 1 more Smart Citation
“…FDR can rely on a well-established calibration procedure, such as the SOL calibration [32]. The error model can be identified using a signal flow graph, and the SOL-calibrated reflection scattering parameter can be calculated from measurements of , , and , which are the reflection scattering parameters when an open, short, and matched loads, respectively, are connected.…”
Section: B Calibration Issues In Fdmentioning
confidence: 99%
“…Measurements at higher frequencies, above about 1 MHz, require extra attention to circuit components, connections, and calibration [10].…”
Section: Instrumentationmentioning
confidence: 99%