The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets—Quantification and possible corrections

Abstract: Purpose To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D98, target volume coverage (TVC), conformity (CI), and gradient index (GI). Methods Winston‐Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assurance (QA) plan was applied to the ArcCHECK phantom to assess the impact of isocenter shift corrections on Gamma pass rates. These corrections included gantry sag, distance of collimator and couch axes… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
17
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 9 publications
(20 citation statements)
references
References 25 publications
2
17
0
Order By: Relevance
“…Combining the corrections is straight forward. A script generated using Pinnacle (Philips Radiation Oncology Systems, Fitchburg, WI, version 16.2) presented by Wack et al 4 creates the true isocenter per beam direction. This is the nominal isocenter shifted due to gantry sag and wobble, the distance between CBCT and MV isocenter, the distance of the collimator axis and couch axis to the gantry axis.…”
Section: Correction Proceduresmentioning
confidence: 99%
See 3 more Smart Citations
“…Combining the corrections is straight forward. A script generated using Pinnacle (Philips Radiation Oncology Systems, Fitchburg, WI, version 16.2) presented by Wack et al 4 creates the true isocenter per beam direction. This is the nominal isocenter shifted due to gantry sag and wobble, the distance between CBCT and MV isocenter, the distance of the collimator axis and couch axis to the gantry axis.…”
Section: Correction Proceduresmentioning
confidence: 99%
“…All planned dose distributions were calculated and exported using (1 mm) 3 voxel size. While Low and Dempsey 15 suggest using a spatial criterion as three times the calculation grid size or more for accurate calculation of gamma values in steep dose gradient regions, here, for the calculation grid size of 1 mm a stricter criterion of 2%/2 mm was chosen, as in Wack et al 4 (see this reference for a further discussion of the choice). The common clinical choice of a gamma pass rate was also evaluated for the 1%/1 mm criterion.…”
Section: Ta B L Ementioning
confidence: 99%
See 2 more Smart Citations
“…RT nominally requires a geometric accuracy of ≤2 mm or 0.5-1 mm for stereotactic radiosurgery (SRS). [21][22][23] The peak-to-peak deviation from the average inhomogeneity reported for the 1.5 T Elekta Unity was measured for a single slice in a 40 cm diameter cylindrical phantom at a repetition rate of 0.7 Hz using field maps. 6 The measured field homogeneity was not affected by gantry rotation speeds of 1 or 2 rpm.…”
Section: Field Homogeneity and Center Frequency Measurementsmentioning
confidence: 99%