The Effects of Multiple Infusion Line Extensions on Occlusion Alarm Function of an Infusion Pump

Abstract: When simulating low flow rate infusions (5 mL/h) as for vasopressor support, occlusion alarm time was critically prolonged, especially with an increased length of infusion lines.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
8
0

Year Published

2014
2014
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 12 publications
(9 citation statements)
references
References 9 publications
0
8
0
Order By: Relevance
“…Testing to maintain PPP within a safe range is also required, where PPP high enough to cause fluid leakage or rupture should not be generated. If a single large volume is injected in the patient as bolus or infusion is blocked, an alarm should sound immediately [ 11 ]. Continuous backflow under normal operating conditions or even with a simple defect can be dangerous, and the device must not allow this to happen under any circumstance.…”
Section: Discussionmentioning
confidence: 99%
“…Testing to maintain PPP within a safe range is also required, where PPP high enough to cause fluid leakage or rupture should not be generated. If a single large volume is injected in the patient as bolus or infusion is blocked, an alarm should sound immediately [ 11 ]. Continuous backflow under normal operating conditions or even with a simple defect can be dangerous, and the device must not allow this to happen under any circumstance.…”
Section: Discussionmentioning
confidence: 99%
“…As reported previously, highest possible infusion rate and smaller sized syringes are useful in decreasing alarm delays [7,8,9]. Routine use of a less compliant syringe and syringe extension line might be recommended to reduce the alarm delay [6,10]. During the infusion of drugs requiring a smaller amount of initial bolus, such as remifentanil, lowering the setting of occlusion alarm threshold (P set ) could decrease the alarm delay.…”
Section: Discussionmentioning
confidence: 99%
“…Detection of occlusion is complicated by a prolonged time to alarm using low flow rates, longer infusion lines, larger syringe size, and varying pump models. 33,34 Other alarms were documented 25 as a warning for low battery (2.9%), a relatively easy fix if pump batteries are kept charged, and proximal air (2.52%).…”
Section: Prevalence and Causes Of Smart Pump Alerts And Alarmsmentioning
confidence: 99%