The Dilemma of ICD Implant Testing

Abstract: Ventricular fibrillation (VF) has been induced at implantable cardioverter defibrillator (ICD) implant to ensure reliable sensing, detection, and defibrillation. Despite its risks, the value was self-evident for early ICDs: failure of defibrillation was common, recipients had a high risk of ventricular tachycardia (VT) or VF, and the only therapy for rapid VT or VF was a shock. Today, failure of defibrillation is rare, the risk of VT/VF is lower in some recipients, antitachycardia pacing is applied for fast VT… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
112
1
10

Year Published

2012
2012
2019
2019

Publication Types

Select...
5
3

Relationship

2
6

Authors

Journals

citations
Cited by 130 publications
(125 citation statements)
references
References 218 publications
2
112
1
10
Order By: Relevance
“…This comparison cannot be assessed statistically due to the small number of unsuccessful shocks. This rate of first shock failure is in agreement to that reported for spontaneous VF in patients with DT (80-91%) or safety margin testing (91%) at ICD implantation [2,4,5]. Similarly, a significant proportion (80-87%) of VT/fast VT episodes were successfully ATP-terminated without shock delivery, regardless of whether DT was performed.…”
Section: Icd Therapy Efficacy In Dt+ Vs Dtpatientssupporting
confidence: 78%
See 2 more Smart Citations
“…This comparison cannot be assessed statistically due to the small number of unsuccessful shocks. This rate of first shock failure is in agreement to that reported for spontaneous VF in patients with DT (80-91%) or safety margin testing (91%) at ICD implantation [2,4,5]. Similarly, a significant proportion (80-87%) of VT/fast VT episodes were successfully ATP-terminated without shock delivery, regardless of whether DT was performed.…”
Section: Icd Therapy Efficacy In Dt+ Vs Dtpatientssupporting
confidence: 78%
“…The assessment of defibrillation efficacy has long been the standard of care during implantable cardioverter-defibrillator (ICD) insertion, and is recommended by ICDs manufacturers and scientific guidelines [1,2]. However, this strategy is currently being questioned due to the high probability of current devices to successfully defibrillate ventricular arrhythmias, the low incidence of spontaneous ventricular fibrillation (VF) (~10%) and the high efficacy of dedicated antitachycardia pacing (ATP) strategies [2][3][4][5][6][7][8].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…The authors concluded that defibrillation efficacy without DT was noninferior to ICD implantation with DT in left-sided ICD implants. Because no major benefit or harm associated with DT was detected, in patients with a leftsided pectoral implantation it is reasonable to omit routine VF induction and DT during ICD implantation, assuming stable ICD lead position and good sensing and capture function [217][218][219][220]. This approach is particularly applicable to patients with ischemic and idiopathic dilated cardiomyopathy, given these entities were well represented in the studied cohort.…”
Section: Iia B-nrmentioning
confidence: 99%
“…High DFT is defined as an absolute value of shock energy >25Joules (J) or a safety margin of <10J below the maximum output of the device. This is assessed by two successful shocks of same strength [8] and the reported incidence of high DFT is from 2 to 24%. Russo et al reported a 6.2% prevalence of high DFT (n=1139) [9] which was replicated in a separate study by Osswald and colleagues in a larger population (n= 2803) [10].…”
Section: Introductionmentioning
confidence: 99%