2010
DOI: 10.1088/0026-1394/47/2/s14
|View full text |Cite
|
Sign up to set email alerts
|

The determination of the uncertainties of spectral emissivity measurements in air at the PTB

Abstract: The instrumentation of the Physikalisch-Technische Bundesanstalt (PTB) for the measurement of the directional spectral emissivity in air in the temperature range from 80 • C to 400 • C and wavelength range from 4 µm to 40 µm and the subsequent procedures for data evaluation are described. Special emphasis is placed on the calculation of the uncertainty of the directional spectral emissivity and of the uncertainty of the sample surface temperature. For samples with typical emissivities of about 0.6, absolute ex… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
40
0
1

Year Published

2010
2010
2020
2020

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 64 publications
(42 citation statements)
references
References 6 publications
1
40
0
1
Order By: Relevance
“…The emissivity measurements of the black coatings were taken at two setups at PTBunder vacuum at the RBCF and in air at the EMAF-by applying the measurement schemes and data, and uncertainty evaluation described shortly above and described in detail in [1][2][3]6,7]. Two sets of detectors and beamsplitters were used with the IR-FTS according to the wavelength and temperature range of the measurements.…”
Section: Experimental Results On Different Black Coatingsmentioning
confidence: 99%
See 3 more Smart Citations
“…The emissivity measurements of the black coatings were taken at two setups at PTBunder vacuum at the RBCF and in air at the EMAF-by applying the measurement schemes and data, and uncertainty evaluation described shortly above and described in detail in [1][2][3]6,7]. Two sets of detectors and beamsplitters were used with the IR-FTS according to the wavelength and temperature range of the measurements.…”
Section: Experimental Results On Different Black Coatingsmentioning
confidence: 99%
“…Highly accurate measurements of directional spectral, directional total and hemispherical total emissivities for four widely used black coatings-Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black-were taken by PTB in the wavelengths range from 4 µm to 100 µm at different temperatures in the range from 25 • C to 150 • C. The achieved absolute standard uncertainty (k = 1) is calculated according to [5][6][7]. It is 0.005 or better for measurements at high temperatures (T = 120 • C and T = 150 • C) and it is better than 0.022 (for some measurements better than 0.011) for measurements at 25 • C. It should be noted that the achieved uncertainty significantly depends on the experimental conditions and the investigated coating.…”
Section: Resultsmentioning
confidence: 99%
See 2 more Smart Citations
“…Furthermore, a second reference blackbody of significantly different temperature from the first reference blackbody is generally applied for the correction of the background signal. The instrumentation and the associated uncertainties of this measurement principle are published in detail in [7][8][9], respectively. To determine the angular dependency of the directional spectral emissivity from a series of the above described ratio measurements the sample can be tilted in a range from 5°to 70°between the sample surface normal and the direction of observation.…”
Section: Setup For Directional Emissivity Measurementmentioning
confidence: 99%