2016
DOI: 10.1117/12.2231415
|View full text |Cite
|
Sign up to set email alerts
|

The design, implementation, and performance of the Atro-H SXS calorimeter array and anti-coincidence detector

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
20
0

Year Published

2017
2017
2018
2018

Publication Types

Select...
8

Relationship

3
5

Authors

Journals

citations
Cited by 21 publications
(20 citation statements)
references
References 12 publications
0
20
0
Order By: Relevance
“…The OVCS, MVCS, and IVCS were enshrouded by 50 layers, 30 layers, and 20 layers of multilayer insulation films, respectively. The JTS was cooled by a 4 He JT cryocooler that used two 2ST cryocooler units as precoolers (PC-A and PC-B), whereas the IVCS and the OVCS were cooled by two 2ST cryocooler units called shield coolers (SC-A and SC-B). The JT cryocooler was equipped with two compressor units in series, that were called JT-CMP-L and JT-CMP-H.…”
Section: Designmentioning
confidence: 99%
See 1 more Smart Citation
“…The OVCS, MVCS, and IVCS were enshrouded by 50 layers, 30 layers, and 20 layers of multilayer insulation films, respectively. The JTS was cooled by a 4 He JT cryocooler that used two 2ST cryocooler units as precoolers (PC-A and PC-B), whereas the IVCS and the OVCS were cooled by two 2ST cryocooler units called shield coolers (SC-A and SC-B). The JT cryocooler was equipped with two compressor units in series, that were called JT-CMP-L and JT-CMP-H.…”
Section: Designmentioning
confidence: 99%
“…1 The soft x-ray spectrometer (SXS) 2,3 was a cryogenic high-resolution x-ray spectrometer onboard Hitomi, utilizing an x-ray microcalorimeter array. 4 Operated at 50 mK, it was designed to achieve a resolving power of 1000 or larger at 6 keV. The detector array was cooled to 50 mK using a three-stage adiabatic demagnetization refrigerator (ADR).…”
Section: Introductionmentioning
confidence: 99%
“…The ADRC also controls the heaters on the optical blocking filters. 8 The heaters for both the IVCS and outer vapor-cooled shields (OVCS) filters were checked for their aliveness on Y þ 13. The temperature control of the Dewar main shell (DMS) filter heater was tested for 3.5 h on Y þ 13.…”
Section: Adiabatic Demagnetization Refrigerator Controllermentioning
confidence: 99%
“…4) in order to aid in planning for future microcalorimeter missions. Details are found in other articles of the same special section for the spacecraft, 1 the overall mission instrument, 2 the sensor and its performance, [8][9][10][11] the mirror and the aperture assembly, 12,13 the cooling chain, 5,14-17 the vibration isolation system, 18 the filter wheel (FW), 4 and the signal processing. 19,20 All data presented here are available in Ref.…”
Section: Introductionmentioning
confidence: 99%
“…The SXS detector array 4 is a 6 × 6 array of square pixels arranged with a 0.832-mm pitch, kept at a temperature of 50 mK, mounted inside a Dewar. The detector array has a minimum energy range of 0.3 to 12 keV and demonstrated an inflight energy resolution of 4.9 eV.…”
mentioning
confidence: 99%