The accuracy of the Acuros XB algorithm in external beam radiotherapy – a comprehensive review

Abstract: Review Article AbstractThe accuracy of dose calculation algorithms has been a topic of interest among the radiotherapy community throughout last decades. On one hand the advancements in computers and algorithms has improved the accuracy, but on the other hand the developments in other parts of treatment process, in treatment delivery techniques and in treatment devices have always pushed the requirements to the next level. In this review article a comprehensive overview on the accuracy of a new type 'c' dose c… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

2
39
0

Year Published

2014
2014
2023
2023

Publication Types

Select...
9
1

Relationship

0
10

Authors

Journals

citations
Cited by 51 publications
(42 citation statements)
references
References 64 publications
(90 reference statements)
2
39
0
Order By: Relevance
“…The methodology applied in this study can also be implemented to validate other advanced dose calculation algorithms such as Acuros XB algorithm and anisotropic analytical algorithm. [25][26][27] Conclusion Dose calculation using Prowess TPS based on CCCS algorithm is generally in excellent agreement with measurements. There are some points with deviation which was not within the limits that have been set up for dose planning systems; the largest deviation was found at the small field size 4 cm × 4 cm.…”
Section: Discussionsupporting
confidence: 62%
“…The methodology applied in this study can also be implemented to validate other advanced dose calculation algorithms such as Acuros XB algorithm and anisotropic analytical algorithm. [25][26][27] Conclusion Dose calculation using Prowess TPS based on CCCS algorithm is generally in excellent agreement with measurements. There are some points with deviation which was not within the limits that have been set up for dose planning systems; the largest deviation was found at the small field size 4 cm × 4 cm.…”
Section: Discussionsupporting
confidence: 62%
“…In the AAA all energy from a photon interaction is deposited either in the forward beam direction or along one of 16 lateral transport lines, all located in the plane perpendicular to the incident beam direction. AAA presents more accurate algorithm as compared to pencil kernel algorithm (8)(9)(10).…”
Section: Dose Calculation Algorithmsmentioning
confidence: 99%
“…17,18,19 Moreover, the choice of the treatment planning systems is very important in radiotherapy dose calculation, especially in the presence of treatment planning complexity such as irregular shaped tumors and tumor heterogeneities. 20,21 Volume …”
Section: Dose Calculation Accuracymentioning