Technical Note: Adapting a GE SIGNA PET/MR scanner for radiotherapy

Abstract: We present a tabletop and coil holder for an anterior array coil to be used with a GE SIGNA PET/MR scanner, for scanning patients in the radiotherapy work flow. Implementing attenuation correction of the added hardware from the radiotherapy setup leads to acceptable PET image quantitation. The drop in SNR in MR images may require adjustment of the imaging protocols.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
14
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 11 publications
(16 citation statements)
references
References 22 publications
2
14
0
Order By: Relevance
“…The increased separation between the coil and the phantom led to a 32.4% drop in SNR. This was within the SNR reductions of 44% [23] , 42% [10] , 33% [9] and 14% [33] reported in literature for similar MRI in RT image acquisitions.…”
Section: Discussionsupporting
confidence: 89%
See 1 more Smart Citation
“…The increased separation between the coil and the phantom led to a 32.4% drop in SNR. This was within the SNR reductions of 44% [23] , 42% [10] , 33% [9] and 14% [33] reported in literature for similar MRI in RT image acquisitions.…”
Section: Discussionsupporting
confidence: 89%
“…RT-specific demands like patient positioning (flat tabletop-inserts) and immobilization devices needed to replicate patient treatment position also affect MR image quality. For example, by using a flat tabletop-insert and coil holder, Brynolfsson et al [9] reported that SNR reduced to 66% whilst Xing et al [10] , had an average SNR decrease of 42%. Batumalai et al [11] , showed that for supine RT-specific imaging of the breast, a switch from the standard diagnostic 16-channel prone coil to an 18-channel surface coil decreased SNR by 41% and 45% for supine flat and supine inclined (10°) respectively.…”
Section: Introductionmentioning
confidence: 99%
“…A few solutions have been proposed for adapting PET/ MRI scanners to meet RT requirements. 29,30 For patients with HNC, Paulus et al 13 proposed an alternative patient setup and coil arrangement that, unlike standard head and neck coil usage, allows proper patient positioning with fixation masks. Winter et al 31 further studied the feasibility and image quality of this setup and concluded that the requirements for RT treatment planning were met.…”
Section: Introductionmentioning
confidence: 99%
“…On the other hand, the MRI cannot be used for intra-fraction monitoring, and additional time is required to move the MRI scanner in and out of the room. A similar approach using a 1.5 T scanner was developed in Umea in which the patient rather than the MRI scanner is moved (Karlsson et al 2009), although this system has recently been decommissioned and replaced with a PET/MRI scanner (Brynolfsson et al 2018). Table II shows a comparison of existing in-room/near-room MRI systems.…”
Section: In-room Mri Radiotherapy Systemsmentioning
confidence: 99%