2023
DOI: 10.1039/d3re00038a
|View full text |Cite
|
Sign up to set email alerts
|

Synthesis of [68Ga]Ga-PSMA-11 using the iMiDEV™ microfluidic platform

Abstract: We report herein the first production of [68Ga]Ga-PSMA-11 based on a microfluidic disposable cassette using iMiDEVTM synthesizer. The main steps of synthesis, 68Ga-concentration on the cation exchange resin followed by...

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
3

Relationship

2
1

Authors

Journals

citations
Cited by 3 publications
(6 citation statements)
references
References 29 publications
0
5
0
Order By: Relevance
“…However, after conducting heating element studies to compare the set value with the actual temperature value inside R2, it became evident that there was a significant variation between the two values. This discrepancy was attributed to the heat transfer between the heating component and the R2 chamber, which was further confirmed through a simulation study (Ovdiichuk et al 2023 ). The actual temperature inside the R2 chamber was approximately 70% of the set value.…”
Section: Discussionmentioning
confidence: 63%
See 2 more Smart Citations
“…However, after conducting heating element studies to compare the set value with the actual temperature value inside R2, it became evident that there was a significant variation between the two values. This discrepancy was attributed to the heat transfer between the heating component and the R2 chamber, which was further confirmed through a simulation study (Ovdiichuk et al 2023 ). The actual temperature inside the R2 chamber was approximately 70% of the set value.…”
Section: Discussionmentioning
confidence: 63%
“…Cassettes with a pressure drop of 0.2 to 0.3 bar were selected to perform trapping and elution on R1. The concentration of [ 68 Ga]GaCl 3 was performed using Chromafix PS-H + resin following the methodology proposed in previous studies (Ovdiichuk et al 2022 , 2023 ), and the elution technique was adapted according to the work of Mueller et al ( 2012 ). In this approach, R1 (50 µL) was filled with Chromafix PS-H + resin to concentrate the [ 68 Ga]GaCl 3 .…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The chip contains a single liquid phase reactor and three solid phase columns that can be accessed at the operator's command via 34 pneumatically actuated microfluidic valves. 101 It also contains a series of vial connections on which preloaded vials of reagents are loaded. iMiDEV™ has been used to produce [ 18 F]NaF in a fully automated fashion in only 8 min with a RCY of 82%.…”
Section: Microfluidic Radiotracer Productionmentioning
confidence: 99%
“…102 It was also shown to automatically synthesize quality 11 C and 68 Ga radiotracers. 101,103,104 This platform represents a major milestone as one of the few fully integrated and customizable microfluidic platforms for radiotracer synthesis.…”
Section: Microfluidic Radiotracer Productionmentioning
confidence: 99%