Survey of mammography practice in Croatia: equipment performance, image quality and dose

Abstract: A national audit of mammography equipment performance, image quality and dose has been conducted in Croatia. Film-processing parameters, optical density (OD), average glandular dose (AGD) to the standard breast, viewing conditions and image quality were examined using TOR(MAM) test object. Average film gradient ranged from 2.6 to 3.7, with a mean of 3.1. Tube voltage used for imaging of the standard 45 mm polymethylmethacrylate phantom ranged from 24 to 34 kV, and OD ranged from 0.75 to 1.94 with a mean of 1.2… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
5
0

Year Published

2009
2009
2022
2022

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 12 publications
(5 citation statements)
references
References 7 publications
0
5
0
Order By: Relevance
“…A to tal of 20 mammographic units of dif fer ent man u fac tur ers were in cluded in the study, such as: Elscint Glory (1 mammographic unit), GE-CGR Senographe 800 T (1), Hologic Lorad M IV (4), Metaltronica Lilyum (2), Philips Mammo Diagnost 3000 (2), hilips Mammo Diagnost BC (1), Philips Mammo Diagnost UC (1), Philips Mammo Diagnost UM (1), Planmed Sophie (4), Siemens Mammomat 1000 (2), and Siemens Mammomat 3000 (1). All the units were based on the screen-film com bi na tion as the im age re cep tor, rou tinely uti lized in clin i cal mam mogra phy.…”
Section: Selection Of Mammographic Unitsmentioning
confidence: 99%
See 1 more Smart Citation
“…A to tal of 20 mammographic units of dif fer ent man u fac tur ers were in cluded in the study, such as: Elscint Glory (1 mammographic unit), GE-CGR Senographe 800 T (1), Hologic Lorad M IV (4), Metaltronica Lilyum (2), Philips Mammo Diagnost 3000 (2), hilips Mammo Diagnost BC (1), Philips Mammo Diagnost UC (1), Philips Mammo Diagnost UM (1), Planmed Sophie (4), Siemens Mammomat 1000 (2), and Siemens Mammomat 3000 (1). All the units were based on the screen-film com bi na tion as the im age re cep tor, rou tinely uti lized in clin i cal mam mogra phy.…”
Section: Selection Of Mammographic Unitsmentioning
confidence: 99%
“…Mam mog ra phy is the method of choice for early de tec tion of breast can cer and a well ac cepted screening method for women of cer tain age groups [1][2][3]. Using mam mog ra phy, it is pos si ble to de tect can cer in a phase when treat ment is most ef fec tive, but only if a con stant pro duc tion of qual ity im ages in terms of contrast and res o lu tion is as sured.…”
Section: Introductionmentioning
confidence: 99%
“…[8][9][10][11][12] The lesson learned is significant variations in patient doses between different radiological departments for the same type of examination.…”
Section: Introductionmentioning
confidence: 99%
“…Therefore, several authors [6][7][8][9][10][11][12][13] proposed that the average X-ray dose to the glandular tissue (MGD) is the most appropriate dosimetric quantity to predict the risk of carcinogenesis. It can be calculated from measurements on patients using breast thickness, breast composition, and X-ray exposure factors.…”
mentioning
confidence: 99%
“…Risk factors for interval breast cancer. The risk factors for developing interval cancer have not been fully explored [8,[10][11][12][13][14][15].…”
mentioning
confidence: 99%