2021
DOI: 10.1016/j.ejmp.2021.04.018
|View full text |Cite
|
Sign up to set email alerts
|

Surgivisio® and O-arm®O2 cone beam CT mobile systems for guidance of lumbar spine surgery: Comparison of patient radiation dose

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
10
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 10 publications
(10 citation statements)
references
References 30 publications
0
10
0
Order By: Relevance
“…However, Rousseau et al performed a dosimetric comparison in experimental conditions. 21 The authors found an effective dose related to 3D of 2.41 and 0.35 mSv with the O-arm and Surgivisio device, respectively, and an effective dose related to 2D of 1.54 and 0.30 mSv, respectively. They also reported that organ doses were 5 to 7 times higher with the O-arm in comparison with the Surgivisio device, which followed a trend that aligned with our results.…”
Section: Discussionmentioning
confidence: 94%
See 1 more Smart Citation
“…However, Rousseau et al performed a dosimetric comparison in experimental conditions. 21 The authors found an effective dose related to 3D of 2.41 and 0.35 mSv with the O-arm and Surgivisio device, respectively, and an effective dose related to 2D of 1.54 and 0.30 mSv, respectively. They also reported that organ doses were 5 to 7 times higher with the O-arm in comparison with the Surgivisio device, which followed a trend that aligned with our results.…”
Section: Discussionmentioning
confidence: 94%
“…Voltage is higher in standard O-arm settings in comparison with the Surgivisio device, leading to higher level of irradiation. 21 Higher effective dose and better image quality in 2D mode for the O-arm can be explained by its settings, with a standard 30 pulses per second. Surgivisio produces 8 pulses per second in 2D mode.…”
Section: Discussionmentioning
confidence: 99%
“…Once the 3D CBCT images of the levels to be instrumented were acquired with the Surgivisio imaging platform [ 35 ], screws were placed using the associated navigation system. Rods were bent to fit the correct expected sagittal balance.…”
Section: Methodsmentioning
confidence: 99%
“…A more tightly integrated system is the recently debuted eCential platform from the French company eCential Robotics, rings together intraoperative 2D/3D imaging, navigation and robotics, preventing pitfalls of coordinate frame registration [63].…”
Section: Figmentioning
confidence: 99%