2006
DOI: 10.1118/1.2240984
|View full text |Cite
|
Sign up to set email alerts
|

SU-FF-T-56: A Simple Method for Selecting a Pinnacle IMRT Point for Verification in RadCalc

Abstract: Purpose: To select a reference point in a low dose gradient region of an IMRT treatment plan to enhance the MU and point dose agreement between Pinnacle and RadCalc. Method and Materials: After generating an IMRT plan within Pinnacle, we export it to RadCalc for a second check of the MU's. Frequently, the MU difference is significant for a plan with split beams or isocenter out of the field. In contrast to Pinnacle, RadCalc displays a coordinate grid over its BEV fluence. By utilizing this feature for the prob… Show more