Abstract:Purpose:
To evaluate the performance of the stealth chamber by IBA under large field conditions.
Methods:
Measurement of PDDs and profiles for fields of 10×10, 15×15, 20×20 and 25×25 cm2 were acquired in a blue phantom2 scanning system for 6 and 15 MV beams in a Varian TrueBeam linear accelerator. The stealth chamber has a field size limit of 25×25 cm2. Measurements with two different detector set were performed. First a cc13 chamber as field and reference chamber was acquired. And with a cc13 chamber as field… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.