2015
DOI: 10.1118/1.4924608
|View full text |Cite
|
Sign up to set email alerts
|

SU‐E‐T‐246: Detector Selection for Small Field Measurements, a Comparative Study

Abstract: Purpose: To compare commercially available detectors for small field dosimetry. Methods: PTW (Freiburg, Germany) diode E, PTW microDiamond, and SunNuclear (Melbourne, FL) Edge detector were used to measure output factors, PDD, and profiles on a Varian iX linear accelerator (Varian Medical Systems, Palo Alto, CA) using 6MV and 18MV photon beams. The water tank was set to 100cm SSD. Centering profiles at 5cm and 20cm depth confirmed an offset <0.1mm. High resolution PDD (0.1mm) aligned the collecting volume at w… Show more

Help me understand this report

This publication either has no citations yet, or we are still processing them

Set email alert for when this publication receives citations?