Abstract:Purpose:
To compare commercially available detectors for small field dosimetry.
Methods:
PTW (Freiburg, Germany) diode E, PTW microDiamond, and SunNuclear (Melbourne, FL) Edge detector were used to measure output factors, PDD, and profiles on a Varian iX linear accelerator (Varian Medical Systems, Palo Alto, CA) using 6MV and 18MV photon beams. The water tank was set to 100cm SSD. Centering profiles at 5cm and 20cm depth confirmed an offset <0.1mm. High resolution PDD (0.1mm) aligned the collecting volume at w… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.