Output Factors and dose profiles measurements performed, respectively, with the PSD and the PSDCA were in agreement with those obtained with the UD and EBT2 films. For stereotactic radiosurgery treatment verification, the PSD gives accurate results compared to the planning system and the IC once the latter is corrected to compensate for the averaging effect of the IC. The PSD provides precise results when used as a single detector or in a dense array, resulting in a great potential for stereotactic radiosurgery QA measurements.