Study of feasibility of portal dosimetry in comparison with ImatriXX 2-D array system for IMRT and Rapid arc patient specific QA.

Abstract: -This study aimed to compare the Varian portal dosimetry system and the ImatriXX 2-D array system of IBA dosimetry for IMRT and Rapid arc patient specific quality assurance (QA). Fifteen cases of IMRT and fifteen cases of Rapid arc plans were selected for the patient specific QA study using portal dosimetry and ImatriXX 2-D array system. Planar dose comparison was carried out with gamma criteria of 3% -3 mm {dose difference and distance-to-agreement (DTA)}. For portal dosimetry system area gamma, average gamma… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2018
2018
2019
2019

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 14 publications
(13 reference statements)
0
2
0
Order By: Relevance
“…AS1200 detector has an active area of 40 Â 40 cm 2 with 1190 Â 1190 pixel arrays and pixel pitch of 0.336 mm which is suitable for both large-and small-field verification. [3][4][5] The treatment planning system is Eclipse v13. 6.…”
Section: Methodsmentioning
confidence: 99%
“…AS1200 detector has an active area of 40 Â 40 cm 2 with 1190 Â 1190 pixel arrays and pixel pitch of 0.336 mm which is suitable for both large-and small-field verification. [3][4][5] The treatment planning system is Eclipse v13. 6.…”
Section: Methodsmentioning
confidence: 99%
“…Although the primary purpose of EPID is to verify patient positioning, but with the introduction of on board imagers for image guidance the EPID has been more utilised for machine QA and pre-treatment patient specific QA [11][12][13][14]. Dosimetry using EPIDs or portal dosimetry has received considerable attention recently due to its relatively high-resolution of 0.392mm amorphous silicon (a-Si) flat-panel detector [15][16][17][18]. Several authors have extensively studied the dosimetric properties of a-Si 500 and a-Si 1000 EPID and its application in IMRT [19][20][21][22][23][24][25].…”
Section: Introductionmentioning
confidence: 99%