Stability Indicating Rp-HPLC Method for Separation and Quantification of Related Substances of Ticlopidine in Bulk and Pharmaceutical Formulations

Abstract: A simple, specific, accurate and stable reverse phase liquid chromatographic method was developed for the simultaneous determination of ticlopidine and its related impurities A and B in bulk drug and tablet dosage forms. The analysis has been performed on XTerra C18 column (250 mm×4.6 mm; 5 µ id) and mobile phase containing of methanol and pH 6.8 phosphate buffer in the ratio of 80:20 (V/V). The detection was carried at 228 nm with a flow rate of 1.0 mL/min. The retention times were found to be 8.9, 5.98 and 4… Show more