ST analysis of the fetal electrocardiogram – Comments on recent experimental data

Abstract: In their paper, Andriessen at al present a validation of fetal ECG analysis and the clinical STAN device in midgestation fetal lambs exposed to 25 minutes of umbilical cord occlusion. The study presents results that contrast remarkably from previously published experimental data which raises a number of questions and comments. The most striking finding of Andriessen et al is the recording of an extremely high number of alarms from the STAN equipment during control conditions when no alarms at all ar… Show more