Abstract:A 39-year-old man with pathologic myopia and a dislocated native lens of the right eye underwent pars plana lensectomy and implantation of an iris-fixated 3-piece MA60MA foldable acrylic intraocular lens with -5.0 diopter power. The patient presented 11 months later with spontaneous haptic disinsertion from the optic. The meniscus (concave-convex) configuration of this iris-fixated lens may have contributed to spontaneous haptic-optic disjunction. This is the first such case reported involving a negative diopt… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.