Soft X-ray ARPES for three-dimensional crystals in the micrometre region

Abstract: An endstation dedicated to angle-resolved photoemission spectroscopy (ARPES) using a soft X-ray microbeam has been developed at the beamline BL25SU of SPring-8. To obtain a high photoemission intensity, this endstation is optimized for measurements under the condition of grazing beam incidence to a sample surface, where the glancing angle is 5° or smaller. A Wolter mirror is used for focusing the soft X-rays. Even at the glancing angle of 5°, the smallest beam spot still having a sufficient photon flux for ARP… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
11
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
7
1

Relationship

2
6

Authors

Journals

citations
Cited by 16 publications
(11 citation statements)
references
References 20 publications
(25 reference statements)
0
11
0
Order By: Relevance
“…Monochromatic Al Kα (1486.7 eV) was used as the incident X-ray. The PEH measurements were performed at the BL25SU in SPring-8. , The base pressure of the measurement chamber was 2.8 × 10 –8 Pa. We used Scienta-Omicron DA30 as an electron analyzer . The electron analyzer had a 2D detector which covered an azimuthal angle of ±15° and a polar angle of ±10°.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Monochromatic Al Kα (1486.7 eV) was used as the incident X-ray. The PEH measurements were performed at the BL25SU in SPring-8. , The base pressure of the measurement chamber was 2.8 × 10 –8 Pa. We used Scienta-Omicron DA30 as an electron analyzer . The electron analyzer had a 2D detector which covered an azimuthal angle of ±15° and a polar angle of ±10°.…”
Section: Methodsmentioning
confidence: 99%
“…31,37 The base pressure of the measurement chamber was 2.8 × 10 −8 Pa. We used Scienta-Omicron DA30 as an electron analyzer. 40 The electron analyzer had a 2D detector which covered an azimuthal angle of ±15°and a polar angle of ±10°. The angle between the direction of the electron analyzer and the incident soft X-ray beam was 90°.…”
Section: ■ Experimental Sectionmentioning
confidence: 99%
“…At the downstream end of branch A, angle-resolved photoemission (ARPES) studies with a micron-sized beam in the soft-X-ray region are performed. 19,20) A detection area of diameter around 5-10 μm has been achieved using Walter mirror optics. 19) A potential difficulty when micron-beam ARPES is performed is that the exact position of the observed area may be lost during sample rotation.…”
Section: Upgrade Arrangement and Setup Of The Experimental Apparatusesmentioning