2018
DOI: 10.1016/j.ejmp.2018.05.023
|View full text |Cite
|
Sign up to set email alerts
|

Shielding disk position in intra-operative electron radiotherapy (IOERT): A Monte Carlo study

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 12 publications
(7 citation statements)
references
References 11 publications
0
7
0
Order By: Relevance
“…Using special applicator or beam shaper device are not possible with IORT TPS [28][29][30][31]. Alhamada et al and Kamomae et al have proposed new shielding disks and tungsten paper for organ at risk protection [32,33]. 3D Dosimetry impact verification of using these new materials is possible only with our methodology whereas a limited materials list is available in IOERT TPS.…”
Section: Discussionmentioning
confidence: 99%
“…Using special applicator or beam shaper device are not possible with IORT TPS [28][29][30][31]. Alhamada et al and Kamomae et al have proposed new shielding disks and tungsten paper for organ at risk protection [32,33]. 3D Dosimetry impact verification of using these new materials is possible only with our methodology whereas a limited materials list is available in IOERT TPS.…”
Section: Discussionmentioning
confidence: 99%
“…Besides, as provided in Table 1, the first disk layer is always constructed from a low atomic number element or composition with a light physical density to ensure minimizing the electron backscattering effect from the upper disk surface. This backscatter dose component can considerably affect the dose distribution uniformity inside the target volume, as established by some studies 14–16,20,28–30 …”
Section: Methodsmentioning
confidence: 97%
“…15 These radioprotection disks are the most widespread ones which are clinically used during breast intraoperative electron radiotherapy to minimize both the received dose by the healthy underlying tissues and the electron backscattered dose originating from the upper disk surface. 20,28 As demonstrated in Table 1, the thickness of these disks does not exceed 1 cm, because higher disk thicknesses may not allow the surgeon to prepare an appropriate flap (remaining the tumor bed after the surgery) upon the disk surface.…”
Section: Employed Ioert Shielding Disksmentioning
confidence: 99%
See 1 more Smart Citation
“…The position of the shielding disk during the operation can be checked by a surgeon with a needle or with in vivo measurements, see section 4.1.3.2 for more detail. Based on MC calculations, the shielding plate should consist of materials that can attenuate the beam almost completely for used energies and reduce the backscatter radiation (Martignano et al 2007, Alhamada et al 2018.…”
Section: Clinical Applications Of Iortmentioning
confidence: 99%