Setting Roller Pump Occlusion With the Transonic HT109 Flowmeter

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2004
2004
2009
2009

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 0 publications
0
1
0
Order By: Relevance
“…To avoid tubing rupture and minimize hemolysis, we adapted the roller pump occlusion setting technique described by Snyder et al 14 and set the pump to 15% nonocclusive confirmed with the use of an external tubing ultrasonic flow probe and meter (HT110 Bypass Flowmeter, Transonic Systems, Ithaca, NY, http://www.transonic.com). We have discovered with the use of a more open occlusion technique and S-65-HL tubing that we do not have to walk the roller pump raceway tubing segment and we rarely observe plasma-free hemoglobin levels >30 mg/dL during long-term support.…”
Section: Blood Propulsion Devicesmentioning
confidence: 99%
“…To avoid tubing rupture and minimize hemolysis, we adapted the roller pump occlusion setting technique described by Snyder et al 14 and set the pump to 15% nonocclusive confirmed with the use of an external tubing ultrasonic flow probe and meter (HT110 Bypass Flowmeter, Transonic Systems, Ithaca, NY, http://www.transonic.com). We have discovered with the use of a more open occlusion technique and S-65-HL tubing that we do not have to walk the roller pump raceway tubing segment and we rarely observe plasma-free hemoglobin levels >30 mg/dL during long-term support.…”
Section: Blood Propulsion Devicesmentioning
confidence: 99%