Abstract:The sensitivity and the spatial resolution of a small-field ∞-Camera system based on a Position Sensitive Photomultiplier Tube (PSPMT) on a tomographic level are examined in this study. A cylindrical Gel-Phantom (d = 40 mm, h = 50 mm) with cylindro-conoidal tubes and capillaries (from 64 to 640 mm 3 in volume) containing water solution of 99m T c is used as a test phantom in the present work. A total of 24 projections covering the full angle region (0 0°3 60 0 ) are obtained with the ∞-Camera system under exam… Show more
“…[SPA04],[POL07],[THA09],[POL09],[MIK09],[RAP11]. The system contains the HAMAMATSU R2486-04 cylindrical PSPMT [HAM98] with a 50 mm field of view and provides an intrinsic spatial resolution of ~ 100 μm.…”