Sensing Lead-Related Complications in Patients With Transvenous Implantable Cardioverter-Defibrillators

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

1
23
1

Year Published

1998
1998
2011
2011

Publication Types

Select...
5
2
2

Relationship

1
8

Authors

Journals

citations
Cited by 42 publications
(25 citation statements)
references
References 20 publications
1
23
1
Order By: Relevance
“…In a prior investigation, ECG measurements with subcutaneous leads were at greater risk for motion and muscular artifact compared to epicardial leads during treadmill exercise in animals. 18 However, in our study, QRS detection with subcutaneous leads and standard signal processing was not impaired during normal daily activity or treadmill exercise as evidenced by positive predictive values >95% and sensitivity values >90% (mean 93.3 ± 5.6%) during Symphony support. The selected location of the subcutaneous leads provided excellent R wave and moderate T wave amplitudes, which minimized the number of false 'missed beats' and 'double triggering' events.…”
Section: Discussioncontrasting
confidence: 47%
See 1 more Smart Citation
“…In a prior investigation, ECG measurements with subcutaneous leads were at greater risk for motion and muscular artifact compared to epicardial leads during treadmill exercise in animals. 18 However, in our study, QRS detection with subcutaneous leads and standard signal processing was not impaired during normal daily activity or treadmill exercise as evidenced by positive predictive values >95% and sensitivity values >90% (mean 93.3 ± 5.6%) during Symphony support. The selected location of the subcutaneous leads provided excellent R wave and moderate T wave amplitudes, which minimized the number of false 'missed beats' and 'double triggering' events.…”
Section: Discussioncontrasting
confidence: 47%
“…14 Postoperative complications, though uncommon, have been reported with transvenous leads and include lead dislodgment, lead fracture, subclavian vein thrombosis, sensing abnormalities, and lead infection. 18,26 Subcutaneous lead systems (a term broadly used to indicate lead positioning in the space external to the pericardium but below the dermis) have been developed for use in cardiac rhythm monitors. 1,9,13 Subcutaneous leads have long been theorized to be an adequate surrogate for surface signals 12 and QRS detection, 5 have been shown to be less susceptible to motion artifacts than surface leads, 3 and have recently been used in a subcutaneous implantable cardioverter-defibrillator.…”
Section: Discussionmentioning
confidence: 99%
“…The latter is often the mode of presentation. Oversensing of noise can also be noted during device interrogation: either from stored data of non-sustained episodes or directly during manipulation of the generator pocket [2,4,6,11,14,15] . Oversensing from pace-sense lead dysfunction has to be distinguished from external noise sources, myopotentials or electrical interference with other leads.…”
Section: Introductionmentioning
confidence: 99%
“…The ICD leads are significantly more complex in design than pacemaker leads and may be more susceptible to lead failure (3)(4)(5)(6)(7)(8)(9). The higher sensitivity used in ICD sense amplifiers makes them more susceptible to oversensing of noise signals than pacemakers.…”
mentioning
confidence: 99%