Scleral-Sutured Intraocular Lens Dislocations Secondary to Eyelet Fractures

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
13
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 17 publications
(13 citation statements)
references
References 27 publications
0
13
0
Order By: Relevance
“…Finally, and perhaps most significant to the discussed IOL platform, is the potential for eyelet fracture with over-tightening as reported by Watane et al 12 In this study, the authors note that the original MX60 and MX60T have been replaced with MX60E and MX60ET, which have lower tensile strength at the eyelet and may result in a higher rate of fracture. Pollman et al compared the eyelet strength of MX60 and MX60E and found that that the MX60 was 67% stronger, though the result was not statistically significant.…”
Section: Discussionmentioning
confidence: 65%
“…Finally, and perhaps most significant to the discussed IOL platform, is the potential for eyelet fracture with over-tightening as reported by Watane et al 12 In this study, the authors note that the original MX60 and MX60T have been replaced with MX60E and MX60ET, which have lower tensile strength at the eyelet and may result in a higher rate of fracture. Pollman et al compared the eyelet strength of MX60 and MX60E and found that that the MX60 was 67% stronger, though the result was not statistically significant.…”
Section: Discussionmentioning
confidence: 65%
“…Fixing the 10-0 polypropylene on the haptic might also be confusing in this position, and an easy error of crossing of the 10-0 polypropylene might waste all time potentially gained. IOL damage was also described when passing through the injector in the presence of the suture with higher risk of disinsertion [ 24 ].…”
Section: Discussionmentioning
confidence: 99%
“…In one report, the eyelet was reported to have broken in 20 cases postoperatively and 5 cases intraoperatively. 13 IOL tilt requiring repositioning was a relatively rare complication, reported in the sutureless fixation techniques from 1% to 4.5%, 37,40 with no cases reported in Gore-Tex technique studies; however, it should be noted that many studies did not report lens tilt outcomes.…”
Section: Transient Complicationsmentioning
confidence: 98%
“…It should also be noted that there have been reports of eyelet breakage when using the MX60/E IOL for Gore-Tex sutured fixation. In one report, the eyelet was reported to have broken in 20 cases postoperatively and 5 cases intraoperatively 13…”
Section: Complications Rates Of Secondary Scleral Fixation Of Iolsmentioning
confidence: 99%