2022
DOI: 10.1155/2022/8886813
|View full text |Cite
|
Sign up to set email alerts
|

Safety of Occlutech Septal Occluder ACCELL Flex II for Transcatheter Closure of Secundum Atrial Septal Defects in Children: A Long-Term Follow-Up

Abstract: Objectives. To assess the long-term safety and efficacy of the Occlutech® ACCELL® Flex II device used for atrial septal defect (ASD) closure. This device differs from the regular device by having two very thin patches that are made of polyethylene terephthalate (PET). These patches enhance faster sealing of the defect. Background. Transcatheter closure has become the method of choice to manage most patients with secundum ASDs. There are different types of devices. The regular Occlutech device used to close an … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
6
0

Year Published

2022
2022
2023
2023

Publication Types

Select...
1
1
1

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(6 citation statements)
references
References 11 publications
0
6
0
Order By: Relevance
“…Transcatheter device closure of young children requires more experience and skill [ 18 ]. According to the hemodynamics, defects in patients with ratio of pulmonary blood flow to systemic blood flow (Qp/Qs) >1.5 and/or dilated right atrium and ventricle are advised to be closed [ 8 , 18 , 24 , 25 , 29 ]. According to morphological features, patients must have adequate rims on echocardiographic evaluations [ 18 ].…”
Section: Reviewmentioning
confidence: 99%
See 4 more Smart Citations
“…Transcatheter device closure of young children requires more experience and skill [ 18 ]. According to the hemodynamics, defects in patients with ratio of pulmonary blood flow to systemic blood flow (Qp/Qs) >1.5 and/or dilated right atrium and ventricle are advised to be closed [ 8 , 18 , 24 , 25 , 29 ]. According to morphological features, patients must have adequate rims on echocardiographic evaluations [ 18 ].…”
Section: Reviewmentioning
confidence: 99%
“…Knop et al’s study found mild mitral valve insufficiency in two of 157 patients under three years of age with secundum ASD who were managed with either Amplatzer Multi-Fenestrated Septal Occluder or Cardi-O-Fix ASD Occluder [ 29 ]. Patients in El-Sisi et al’s study experienced the widest variety of valvular problems compared to the rest of the studies [ 24 ]. Among 30 patients aged 5-18 years who underwent transcatheter closure of secundum ASD using Occlutech ACCELL Flex II device, mild mitral and aortic regurgitation were seen in two children each that persisted till the last follow-up, at the time of which 11 more had mild tricuspid regurgitation and five more had mild pulmonary regurgitation [ 24 ].…”
Section: Reviewmentioning
confidence: 99%
See 3 more Smart Citations