2007
DOI: 10.1109/tbme.2006.890145
|View full text |Cite
|
Sign up to set email alerts
|

Safety of Active Implantable Devices During MRI Examinations: A Finite Element Analysis of an Implantable Pump

Abstract: Abstract-The goal of this study was to propose a general numerical analysis methodology to evaluate the magnetic resonance imaging (MRI)-safety of active implants. Numerical models based on the finite element (FE) technique were used to estimate if the normal operation of an active device was altered during MRI imaging. An active implanted pump was chosen to illustrate the method. A set of controlled experiments were proposed and performed to validate the numerical model. The calculated induced voltages in the… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
7
0

Year Published

2007
2007
2022
2022

Publication Types

Select...
7
2

Relationship

2
7

Authors

Journals

citations
Cited by 15 publications
(7 citation statements)
references
References 18 publications
0
7
0
Order By: Relevance
“…The MedStream pump design was refined by finite element modeling to minimize interferences with MRI (26). The pump lacks magnetic rotors or rotating parts that are susceptible to MRI magnetic fields.…”
Section: Methodsmentioning
confidence: 99%
“…The MedStream pump design was refined by finite element modeling to minimize interferences with MRI (26). The pump lacks magnetic rotors or rotating parts that are susceptible to MRI magnetic fields.…”
Section: Methodsmentioning
confidence: 99%
“…The remaining drug volume within the reservoir is detected by a fill level sensor, which determines the volume of the reservoir chamber and converts the calculation to determine the remaining drug volume. Importantly, prior published work from our laboratory (10), has demonstrated the use of finite element modeling (ANSYS Emag 9, Canonsburg, PA, USA) to limit susceptibility of the pump electronics to induced voltages created by MRI‐related electro‐magnetic fields.…”
Section: Methodsmentioning
confidence: 99%
“…For the magneto-static problem, for the interaction of the field with complex geometries, the most used method is the FEM. For the gradient field for which the frequency is a few kHz, approximations can be made on the Maxwell equations and the most common method used is the FEM [7].…”
Section: Methodsmentioning
confidence: 99%