2016
DOI: 10.1186/s13036-016-0034-3
|View full text |Cite
|
Sign up to set email alerts
|

Respiration activity monitoring system for any individual well of a 48-well microtiter plate

Abstract: BackgroundSmall-scale micro-bioreactors have become the cultivation vessel of choice during the first steps of bioprocess development. They combine high cultivation throughput with enhanced cost efficiency per cultivation. To gain the most possible information in the early phases of process development, online monitoring of important process parameters is highly advantageous. One of these important process parameters is the oxygen transfer rate (OTR). Measurement of the OTR, however, is only available for smal… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
81
0

Year Published

2016
2016
2023
2023

Publication Types

Select...
6

Relationship

4
2

Authors

Journals

citations
Cited by 44 publications
(83 citation statements)
references
References 41 publications
0
81
0
Order By: Relevance
“…Therefore, oxygen transfer rates (OTRs) were calculated from the decrease in the oxygen partial pressure in the headspace of the reaction vessel. For determination of the OTR in 48‐round‐well plates, the µRAMOS technique was applied . By integration of OTRs over time, the total oxygen transferred (OT) was calculated.…”
Section: Methodsmentioning
confidence: 99%
“…Therefore, oxygen transfer rates (OTRs) were calculated from the decrease in the oxygen partial pressure in the headspace of the reaction vessel. For determination of the OTR in 48‐round‐well plates, the µRAMOS technique was applied . By integration of OTRs over time, the total oxygen transferred (OT) was calculated.…”
Section: Methodsmentioning
confidence: 99%
“…In this work, a new device for online monitoring in MTPs is presented. The system combines the benefits of BioLector [9] and µRAMOS technology [15]. Because all measurements are based on optical signals, a special measuring sequence had to be developed to avoid interference among various measurement systems.…”
Section: Discussionmentioning
confidence: 99%
“…A 48 round deep-well MTP (MTP-R48-B, m2p-labs GmbH, Baesweiler, Germany) is placed on the shaker table of an in-house constructed orbital shaking machine (shaking diameter d 0  = 3 mm). For OTR measurements in each well, the µRAMOS cover was fixed on top of the MTP [15]. A sterile barrier (900371-T, HJ-Bioanalytik, Erkelenz, Germany) is clamped between µRAMOS cover and MTP to avoid contaminations (Fig.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations