2014
DOI: 10.1128/jcm.00356-14
|View full text |Cite
|
Sign up to set email alerts
|

Reply to “Technical Notes on the Correct Configuration of the Alfred 60/AST Device for the Detection of Urinary Tract Infections”

Abstract: Spezzotti has identified several issues that may have resulted in an underestimation of the performance of the Alfred 60/AST device.First, Dr. Spezzotti questions the authenticity of one of the nine false-negative samples (MB12216254), based on the Alfred 60/ AST device flagging this sample as "invalid." Although this is correct, we followed the manual's recommendations (see page 66 in reference 2), which currently does not suggest excluding such a sample. Furthermore, additional samples which gave the correct… Show more