Repeatability and reproducibility of vessel density measurements on optical coherence tomography angiography in diabetic retinopathy

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

3
11
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
9

Relationship

2
7

Authors

Journals

citations
Cited by 19 publications
(16 citation statements)
references
References 19 publications
3
11
0
Order By: Relevance
“…For example, when macular vessel density, including both macular vessel area and skeletonized vessel density, has been calculated using three different types of OCTA devices, RTVue XR Avanti, Cirrus HD-OCT 5000 and Plex Elite 9000, significant variability in the calculations was found across the devices suggesting that the values of vessel density cannot be easily compared in research studies if the same OCTA device is not used. 21 , 36 , 39 In addition to these fully automated algorithms by devices, other studies have found similar results with custom semiautomated algorithms. 40 …”
Section: Methods To Calculate Macular Vessel Densitymentioning
confidence: 52%
See 1 more Smart Citation
“…For example, when macular vessel density, including both macular vessel area and skeletonized vessel density, has been calculated using three different types of OCTA devices, RTVue XR Avanti, Cirrus HD-OCT 5000 and Plex Elite 9000, significant variability in the calculations was found across the devices suggesting that the values of vessel density cannot be easily compared in research studies if the same OCTA device is not used. 21 , 36 , 39 In addition to these fully automated algorithms by devices, other studies have found similar results with custom semiautomated algorithms. 40 …”
Section: Methods To Calculate Macular Vessel Densitymentioning
confidence: 52%
“… 50–54 , 76 Additionally, many different OCTA devices exist and macular vessel density values from different devices are not comparable. 21 , 22 , 39 Therefore, it is generally recommended that the same type of device be used for longitudinal follow-up of vessel density in the same patient. Also, since not all clinical sites have the same machine, this calculation has not yet seen widespread adoption as a metric in clinical trials and thus for clinical use.…”
Section: Limitations and Future Directions Of Macular Vessel Density Measurements In Diabetic Retinopathymentioning
confidence: 99%
“…First, only normal eyes were examined. Our group has initially assessed the repeatability of OCTA measurements in diabetic eyes in one published separate study 48 —in future, it would be valuable to complete similar analysis with manual quantification using a range of binarization techniques. In addition, three devices were used in this study.…”
Section: Discussionmentioning
confidence: 99%
“…This study has a few limitations. The cohort size was small and of healthy and young individuals, which may not be generalisable to older patients and individuals with diseased eyes, where reduced reproducibility has been reported for 3 × 3 OCTA scans 38 . Widefield 15 × 9-mm 2 scanning protocol was introduced recently and the repeatability study in older patients free from eye diseases is ongoing.…”
Section: Analysis Of Choriocapillaris Flow Void Repeatability the Repeatability Of Choriocapillaris Flow Voidsmentioning
confidence: 99%