2017
DOI: 10.1016/j.clae.2017.03.007
|View full text |Cite
|
Sign up to set email alerts
|

Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

9
58
0
3

Year Published

2017
2017
2024
2024

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 68 publications
(74 citation statements)
references
References 13 publications
9
58
0
3
Order By: Relevance
“…This is close to the results of two recent studies, which report that the Pentacam AXL returns readings that, respectively, differ by a mean of 0.05 mm relative to the IOL Master 700 [9] and 0.026 mm relative to the IOL Master 500 [8]. Such a small difference should not influence IOL power calculation.…”
Section: Discussionsupporting
confidence: 89%
See 2 more Smart Citations
“…This is close to the results of two recent studies, which report that the Pentacam AXL returns readings that, respectively, differ by a mean of 0.05 mm relative to the IOL Master 700 [9] and 0.026 mm relative to the IOL Master 500 [8]. Such a small difference should not influence IOL power calculation.…”
Section: Discussionsupporting
confidence: 89%
“…Shajari compared the Pentacam AXL with the IOL Master 500 and the IOL Master 700 and found no statistically significant differences in AL, keratometry and ACD measurements performed with the three devices [8]. Sel comparing the Pentacam AXL with the IOL Master 700 found significant differences for keratometry values, AL and ACD measurements [9].…”
Section: Introductionmentioning
confidence: 98%
See 1 more Smart Citation
“…Other optical techniques including partial coherence interferometry, optical coherence tomography, laser Doppler interferometry and optical low coherence reflectometry have also become popular, driven by many factors such as the importance of accurate measurement of ocular dimensions for cataract surgery and necessity in myopia research (Table ). Given the various measurement techniques and devices, investigations revealed good repeatability and reproducibility and good agreement between the devices . At the time of this review, devices using laser Doppler interferometry technology have not become commercially available.…”
Section: Methods Of Measuring Axial Elmentioning
confidence: 99%
“…[1][2][3][4][5][6][7] The latest studies published, demonstrated a perfect correlation of AL measurements performed with IOL Master 500, IOL Master 700 and Pentacam AXL, as well as a high precision of AL, ACD and corneal curvature. 8,9 The main part of our pre-cataract screening routine is focused on objective assessment of the ocular surface, the cornea, anterior chamber and crystal lens conditions. Modern tomographers, like the Pentacam, support us in detecting forme fruste keratoconus (FFKC), past refractive surgery and corneal diseases such as Fuchs endothelial dystrophy, or signs of dry eye prior to cataract surgery.…”
mentioning
confidence: 99%