2021
DOI: 10.1080/1091367x.2020.1865966
|View full text |Cite
|
Sign up to set email alerts
|

Reliability and Validity of the Lactate Pro 2 Analyzer

Abstract: This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L −1 for blood lactate (BLa) ≤4.0 mmo… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

1
16
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
9

Relationship

1
8

Authors

Journals

citations
Cited by 23 publications
(17 citation statements)
references
References 15 publications
1
16
0
Order By: Relevance
“…For the purposes of comparing the agreement between ergometers, acceptable 95% LoAs were set at 13 W for load at D max , 15 3 beats•min −1 for HR at D max , 2 mL•kg −1 •min −1 for VO 2 at D max , and 0.4 mmol•L −1 for BLa at D max . 16 These criterion LoA for assessed variables were derived from published test-retest reliability for Functional Threshold Power in cyclists, 15 and interdevice agreement between Lactate Pro 2 analyzers during a parallel incremental testing study in the same cohort of rowers. 16…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…For the purposes of comparing the agreement between ergometers, acceptable 95% LoAs were set at 13 W for load at D max , 15 3 beats•min −1 for HR at D max , 2 mL•kg −1 •min −1 for VO 2 at D max , and 0.4 mmol•L −1 for BLa at D max . 16 These criterion LoA for assessed variables were derived from published test-retest reliability for Functional Threshold Power in cyclists, 15 and interdevice agreement between Lactate Pro 2 analyzers during a parallel incremental testing study in the same cohort of rowers. 16…”
Section: Discussionmentioning
confidence: 99%
“…16 These criterion LoA for assessed variables were derived from published test-retest reliability for Functional Threshold Power in cyclists, 15 and interdevice agreement between Lactate Pro 2 analyzers during a parallel incremental testing study in the same cohort of rowers. 16…”
Section: Discussionmentioning
confidence: 99%
“…Blood lactate concentration was measured at rest and at 1 min-walking intervals in between the exercise bouts. Blood samples were taken from the left jugular vein and whole blood lactate concentration was immediately determined by a hand-held lactate analyser (Lactate Pro 2, Axon Lab AG ® ) ( Arratibel-Lmaz et al, 2017 ; Crotty et al, 2021 ).…”
Section: Methodsmentioning
confidence: 99%
“…Lactate concentrations were measured in a capillary blood sample obtained from participants’ fingertips using a portable analyzer (Lactate 2, Arkray, Kyoto, Japan) [ 31 , 32 ].…”
Section: Methodsmentioning
confidence: 99%