Reducing MRI RF-induced heating for the external fixation using capacitive structures

Abstract: This paper presents a generic method to reduce the radiofrequency (RF) induced heating of external fixation devices during the magnetic resonance imaging (MRI) procedure. A simplified equivalent circuit model was developed to illustrate the interaction between the external fixation device and the MRI RF field. Carefully designed mechanical structures, which utilize capacitive reactance from the circuit model, were applied to the external fixation device to mitigate the coupling between the external fixation de… Show more