1975
DOI: 10.1093/bja/47.12.1239
|View full text |Cite
|
Sign up to set email alerts
|

Rebreathing in a T-Piece: Volunteer and Theoretical Studies of the Jackson-Rees Modification of Ayre's T-Piece During Spontaneous Respiration

Abstract: The Jackson-Rees modification of Ayre's T-piece was examined experimentally and theoretically to see what effects the theoretical restrictions of Onchi, Hayashi and Ueyama (1957) (1:1 I:E ratio, zero deadspace volume) and the square-wave approximation of Mapleson (1958) may have had on the calculated critical fresh-gas flow rate necessary to prevent rebreathing of exhaled alveolar gases. In the experimental examination six conscious alert volunteers breathed through the system as the fresh-gas flow rate was va… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
14
0

Year Published

1978
1978
2018
2018

Publication Types

Select...
5
4
1

Relationship

0
10

Authors

Journals

citations
Cited by 49 publications
(16 citation statements)
references
References 4 publications
2
14
0
Order By: Relevance
“…In the Mapleson D system the VD/VT ratio has been considered to have less influence on the degree of rebreathing. The small effects noted in this study support this view and are in line with the results presented by Willis et al (17).…”
Section: Discussionsupporting
confidence: 94%
“…In the Mapleson D system the VD/VT ratio has been considered to have less influence on the degree of rebreathing. The small effects noted in this study support this view and are in line with the results presented by Willis et al (17).…”
Section: Discussionsupporting
confidence: 94%
“…The high fresh gas flow used in our method not only ensures that expired gas does not escape past leaks at the rim of the mask (Evans et al, 1977) but also ensures that rebreathing of expired gas does not occur. At gas flows of less than twice the minute volume, rebreathing is likely to occur (Willis et al, 1975). Assuming, in the neonate, a minute volume of 300 ml/kg per min the fresh gas flow should therefore exceed 600 ml/kg per min.…”
Section: Discussionmentioning
confidence: 99%
“…In Method 2, we produced a Mapleson F system by modifying the Mapleson E in Method 1 5,6 . Using the reservoir bag from the circle system, we cut off the tip of the bag, and attached it to the 'machine' end of the expiratory tube, distal to the microbial filter (Figures 2A and 2B).…”
Section: Methodsmentioning