2021
DOI: 10.1016/j.avsg.2020.05.006
|View full text |Cite
|
Sign up to set email alerts
|

Real-World Usage of the WavelinQ EndoAVF System

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
14
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 20 publications
(14 citation statements)
references
References 14 publications
0
14
0
Order By: Relevance
“…The rates of hemorrhage and pseudoaneurysms were low, as reported in prior EndoAVF studies. 5,9,22,25 Patency and functionality of the fistulae were satisfactory, with comparable results to traditional open surgical AVFs. 3,11,12,20,22,[29][30][31][32] EndoAVF were often associated with split outflow through more than one superficial vein, providing more options and longer lengths for hemodialysis cannulation.…”
Section: Discussionmentioning
confidence: 79%
See 2 more Smart Citations
“…The rates of hemorrhage and pseudoaneurysms were low, as reported in prior EndoAVF studies. 5,9,22,25 Patency and functionality of the fistulae were satisfactory, with comparable results to traditional open surgical AVFs. 3,11,12,20,22,[29][30][31][32] EndoAVF were often associated with split outflow through more than one superficial vein, providing more options and longer lengths for hemodialysis cannulation.…”
Section: Discussionmentioning
confidence: 79%
“…Percutaneous methods for creation of an autogenous AVF were developed over the last decade to potentially reduce some of the complications of open surgical AVF construction. 3,5,7,9,[20][21][22][23][24][25] Patient preference for percutaneous options was found to be important in the subgroup of patients who were averse to surgical AVF and in those with anatomical restrictions to surgical AVF creation. 26,27 However, the challenges to a percutaneous AVF procedure are many.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…3 If the fistula does not achieve this target flow rate, additional endovascular interventions may be considered to assist maturation including coil embolization of a competing brachial vein or angioplasty of the fistula "anastomosis" itself (Ellipsys fistulas only), or even surgical correction. 5 Doppler sonography is also important in assessing the venous outflow. EndoAVFs are multioutflow fistulas which can drain via the cephalic vein, basilic vein, or both.…”
Section: Fistula Maturationmentioning
confidence: 99%
“…These include hematoma (9%), pseudoaneurysm (3%), arterial dissection (1%), and thrombosis (9%). 5,6 Overall, the complication rate is as low as 5.46%, with most complications correctable through minor procedures. 7 It is worth noting, however, that the relatively complex fistula anatomy relative to traditional sAVF can make ligation or surgical revision very challenging.…”
Section: Complicationsmentioning