Rapid radiographic film calibration for IMRT verification using automated MLC fields

Abstract: A method for measuring a film sensitometric curve using a single sheet of film exposed with a two field step-and-shoot MLC treatment was developed and tested with Kodak XV2 and EDR2 films. With this technique a film sensitometric curve can be completed in only 10 minutes, making it practical to generate new film calibrations daily. This method is applicable to film calibrations for all purposes, but is particularly useful in IMRT treatment verification due to the method's use of small fields. This method agree… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
59
0
1

Year Published

2006
2006
2022
2022

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 69 publications
(62 citation statements)
references
References 6 publications
2
59
0
1
Order By: Relevance
“…Digitizer error can be due to random readout error and systematic response error. The systematic error may be film‐dependent, (1) but can be controlled by periodic recalibration or correction factor measurement. Spectral matching error can be minimized by measurement matching or spectral filtering (4) …”
Section: Discussionmentioning
confidence: 99%
See 4 more Smart Citations
“…Digitizer error can be due to random readout error and systematic response error. The systematic error may be film‐dependent, (1) but can be controlled by periodic recalibration or correction factor measurement. Spectral matching error can be minimized by measurement matching or spectral filtering (4) …”
Section: Discussionmentioning
confidence: 99%
“…The right side was irradiated similarly, except the initial irradiation of all four squares received a greater dose. The irradiation sequence was very similar to that followed by Childress et al (1) except that here the jaws as well as MLC were used to block out the squares that had already received the planned dose. The four‐field, two‐ film exposures were performed similarly (Fig.…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations